Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed230017822850000830000230017801230017818

BrightSpeed HPower, Velocity or Delight HPower

Symptom

Collimator Cal fails in fastcals. Collimator Cal fails in fastcals

Error message

Error: 230017822 ACCS CCal: Collimator calibration failed. See long description for full error Thu Jun 5 07:21:16 2008 Error: 230017822 Host: ct01 Process: accs File: collSweepProcessing.c Line: 1417 ACCS CCal: Collimator calibration failed. The measured aperture 3.850000 exceeds the upper limit of 3.830000 for this mode. The database will not be updated

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Error: 230017818 ACCS CCal: WARNING: Operating point maybe too near the ZRATIO Limits See Long Description for detail Tue Feb 12 10:07:16 2008 Error: 230017818 Host: ct_norwalk1 Process: accs File: collSweepProcessing.c Line: 875 ACCS CCal: WARNING: Operating point maybe too near the ZRATIO Limits

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required