GE Healthcare · CTBrightSpeed26010009050234
BrightSpeed Select or Delight Select
Symptom
"Scan hardware Stopped Scan" with PDU errors "Scan hardware Stopped Scan" with PDU errorss
Error message
Error: 260100090 The Kss Contactor Status Signal changed state unexpectedly. Contactor Status: Open See long description for details Tue Mar 11 20:26:37 2008 Error: 260100090 Host: tgp Process: tgp subsys File: scIsrService.c Line: 153 Tue Mar 11 20:26:37 2008 Host : Table Gantry Processor Ermes # : 260100090 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 153 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Axial Scan: 50234/1/2 PDU KSS Contactor Status. The Kss Contactor Status Signal changed state unexpectedly. Contactor Status: Open
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2351491 14-25A CONTACTOR - NGPDU Kss contactor See long Description Part #: 2351491 Part Type: GP CONSUMABLES Part Weight: 1.32 LBR Repairable: No Returnable: No Hazardous Material: N Description: 14-25A CONTACTOR FRU: Yes Status: Orderable Comments: FLS - 14-25A CONTACTOR. ADDED PER NGPDU RSPL FROM ASIA. MS20544 ON 1/2/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required