GE Healthcare · CTLightSpeedVCT2601329132446531163411915
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Attention The detector is not at the proper temperature. see long description for more detail. Attention The detector is not at the proper temperature. If you wish to proceed with this scan select the "OK" button, however IQ may be degraded including image artifacts. Please contact GE Service
Error message
Error: 260132913 Plenum Plenum Filtered Average Temperature Out of Tolerance See long description for full error Error: 260132913 Plenum Plenum Filtered Average Temperature Out of Tolerance Wed Jan 18 19:50:57 2006 Error: 260132913 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2381 Wed Jan 18 19:50:50 2006 Host : Das Ermes # : 260132913 Exception Class : Secondary Severity : Pri/Soft File : ThermalManager.cxx Line# : 2381 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Plenum Plenum Filtered Average Temperature Out of Tolerance Sensor Temperatures: Left: 2850, Center: 2750, Right: 2850 Filtered Average Temperature: 2538 Limits: 2600, 3400 Thu Jan 19 22:49:08 2006 Error: 260132913 Host: orp Process: orp_subsys File: ThermalManager.cxx Line: 2381 Thu Jan 19 22:49:06 2006 Host : Das Ermes # : 260132913 Exception Class : Secondary Severity : Pri/Soft File : ThermalManager.cxx Line# : 2381 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Plenum Plenum Filtered Average Temperature Out of Tolerance Sensor Temperatures: Left: 3000, Center: 2850, Right: 2750 Filtered Average Temperature: 2555 Limits: 2600, 3400 This error
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.