Pular para o conteúdo principal
GE Healthcare · CTHiSpeed200900025

Patient Schedule fails to display some patients when updated.

Symptom

Patient Schedule fails to display some patients when updated. Patient schedule fails to display some patients when updated.

Error message

Error: 200900025 Parse Failed : Invalid Worklist received from remote HIS(xxxx)(xxxx) :-xxx Note in the error example above that the error is referring to dicom tag number 0010 2154 . The numbers shown in this location of an error message like this will have the leading zeroes stripped off on tags that have leading zeroes such as the example shown. Thu Sep 28 00:01:23 2006 Error: 200900025 Host: gect1 Process: WLServer File: WLapiSystem.c Line: 191 Function: Data Management : OC Processing Patient Worklist Manager error Source File: WLS_mfm.c, Line: 1465, Status: 23 Parse Failed : Invalid Worklist received from remote HIS (10)(2154) :-113 Tue Sep 13 18:13:04 2011 Error: 200900025 Host: LS16 Process: WLServer File: WLapiSystem.c Line: 191 Function: Data Management : OC Processing Patient Worklist Manager error. Source File: WLS_mfm.c, Line: 1471, Status: 23 Parse Failed : Invalid Worklist received from remote HIS(32)(1060) :-114

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
303.1_H3M4 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 303.1_H3M4 303.1_H3M4 for LightSpeed 3.x (H2 Gantry with MDAS) Obsoleted by FMI 25313 which in turn was Obsoleted by FMI 25331 303.1_H3M4 for LightSpeed 3.x (H2 Gantry with MDAS) Obsoleted by FMI 25313 which in turn was Obsoleted by FMI 25331 Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2305092-5 303 Ver 303H3.0M4 2305093-5 303 Ver 303 H3.0M4 2305091-5 2305090-5 Ver 303H3.0 M4 Apps ver 303.1_H3M4.1 FW08-02 Obsoleted by FMI 25313
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.