Pular para o conteúdo principal
GE Healthcare · CTLightSpeed245200230015005230015016

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan with Failure reading cal file information from cal file

Error message

Error: 245200 The current scan has been aborted due to an invalid tracking calibration Wed Jan 18 13:09:37 2006 Error: 245200 Host: cmh_ct_ls1 Process: scanmgr File: scanmgr_lib.cxx Line: 485 The current scan has been aborted due to an invalid tracking calibration

Error: 230015005 Unable to retrieve the requested cal module from the Cal Database Sun Jun 19 21:05:13 2005 Error: 230015005 Host: wodct03403 Process: scanmgr File: cal_db_api.c Line: 451 Function: CAL Processing : SBC Processing Unable to retrieve the requested cal module from the Cal Database

Error: 230015016 Failure reading cal file information from cal file. File name then listed See Long Description for examples. Wed Jan 18 13:09:37 2006 Error: 230015016 Host: cmh_ct_ls1 Process: scanmgr File: cal_db_api.c Line: 1326 Function: CAL Processing : SBC Processing Failure reading cal file information from cal file. File: /usr/g/cal_db/tracking/misc/tracking_constants Sun Sep 24 15:33:17 2006 Error: 230015016 Host: wvct Process: scanmgr File: cal_db_api.c Line: 1326 Function: CAL Processing : SBC Processing Failure reading cal file information from cal file. File: /usr/g/cal_db/tracking/das_gain/das_gain Wed Feb 20 16:26:

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues