Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2601000586185961971

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260100058 Number of offsets acquired does NOT match estimate of number of offsets generated See Long Description for complete error code Error: 260100058 Number of offsets acquired does NOT match estimate of number of offsets generated Host : Table Gantry Processor Ermes # : 260100058 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 625 Function : DATA ACQUISITION : Scan Data Acquisition Scan Type : Helical Scan: 3474/7/1 Number of offsets acquired does NOT match estimate of number of offsets generated HW ERROR: Possible hardware problem. Offsets generated (estimate):61859 Offsets Acquired: (OC):61971 Difference Between Generated and Acquired:112 Test Margin: 45 Triggers

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues