GE Healthcare · CTHiSpeed21000200000446200002730527288200000411
"importimage: cannot reserve space.
Symptom
Popup box comes up reading as follows: "importimage: cannot reserve space. Check available disk space and error logs." See Long Description for details System still works but runs slower than normal. Customer sees a message p opup box coming up continually which reads as follows "importimage: cannot reserve space. Check available disk space and error logs." System still works but runs slower than normal.
Error message
Error: 2113 request exceeds database quota of 21000 images See Long Description for complete error Thu Feb 22 14:45:53 2007 2113 ct01 imserver imsprim.c 678 request exceeds database quota of 21000 images. Error: 200000446 SdC response envelope does not contain a list See long description for full error Wed May 13 07:27:20 2009 Error: 200000446 Host: ch-rad-ct-ge Process: dms File: SdCDataPackage.cxx Line: 697 SdC response envelope does not contain a list. It has a report which means it was not able to process our request. It returned the following error message : [IMServer]> request exceeds database quota of 21000 images Error: 200002730 Image space allocate response size error. See long description for details Sun Jun 3 10:48:22 2007 Error: 200002730 Host: ct01 Process: imagecreate File: ic_imagecreate_req.c Line: 1284 Image : Image space allocate response size error Requested size : 527288 Response path count : 1 Response status : -1 Error: 200000411 Image space allocation failed Tue Apr 10 06:22:01 2007 Error: 200000411 Host: arls Process: dms AllocSpace.cxx Line: 212 Image space allocation failed.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required