Symptom

Hardware reset fails with ORP to (CCB, DCB or Jedi)communication errors. Hardware reset fails with ORP to CCB communication errors.

Error message

Error: 260101008 The ORP failed to connect with the CCB sub-system in the allotted time. Check to see if subsystem is alive.Timeout value: 4000ms error 260101008 The ORP failed to connect with the CCB sub-system in the allotted time. Check to see if subsystem is alive.Timeout value: 4000ms Open LD to see error example Error: 260101008 The ORP failed to connect with the CCB sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS This was followed by Error: 244540 Immediate Request cannot be honored since the firmware is in the DOWN state.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.