GE Healthcare · CTHiSpeedLightSpeed18220710494706366543018220866531
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Cradle will move a few inches and stop. Cradle will move in a few inches and stop
Error message
Error: 182207 HW error: Alignment error between Encoder and Potentiometer See Long Description for complete error Tue Aug 5 15:05:39 1997 Error: 182207 Host: ETC Proc:Cradle File: cmcCorrErr.c 1.22 3/11/92 Line: 51 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:39:320 Log Series: 28 HW error: Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) Error: 182208 Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) See the LD for the complete error code Error: 182208 Alignment error between Encoder and Potentiometer. Tue Aug 5 15:05:36 1997 Suite: PCT2 Host: ETC Proc:Cradle Error: 182208 File: cmcEncoderRef.c 1.8 8/27/ Method: No Method Line: 152 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:35:935 Log Series: 27 HW error: Encoder C Pulse Alignment error with current Potentiometer reading. Current Position 66531 (.01 mm). Encoder Alignment Minimum (14
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-297036G2 longitudinal potentiometer See LD for more info 46-297036G2 longitudinal potentiometer ( NOTE the pot/encoder assy is # 46-296234G1 ) Part #: 46-297036G2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: POT ASSY. FRU: Yes Status: Orderable Comments: FLS - PF8 TWICE FOR FE ASSIST COMMENTS...... COLLECTION OF DEFECTIVES START7/23/01 MS5277 SPP CODE CHANGED FROM 2 TO N PER PCN 220431. MSA8429 03/25/02 PF8 AGAIN FOR FA ASSIST COMMENTS. FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. FLS - FE ASSIST (971512) (MS6588) 46-297036G2 IS THE HSA POTENTIOMETER ASSEMBLY WHICH CONSISTS OF A BURNS 5K POT AND A 2.2 MICRO FARAD CAP. THE CAPACITOR IS SOLDER ACROSS THE WIPER AND RETURN TERMINALS. THE 9800 TABLE ELEVATION POT (46-165135P3) IS THE SAME 5K BURNS POT BUT DOES NOT HAVETHE NOISE REDUCING CAPACITOR PLEASE REFER TO SN-T1336 FOR DETAILS REGARDING CAPACITOR AND POT NOISE DO YOU STILL WISH TO ORDER THIS PART?
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required