GE Healthcare · CTLightSpeed182408

LightSpeed 4.X - GRE

Symptom

Table will not move up, especially when in full down position. See long description Table will not raise or move up, especially when it's in the full down position and this could be intermittent along with noises noticed by the customer.

Error message

Error: 182408 Velocity tracking error. Probable Problem could be 180V Power is LOW. See Long Description for complete error Error: 182408 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 83 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 11:57:28:685 Log Series: 35 HW/SW error: Velocity tracking error. Actual Velocity 0 (.1 mm/sec) Expected Velocity 400 (.1 mm/sec) Tol 320 (.1 mm/sec) Cur Pos 3276 (.1 mm) Elev Direction: 1 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2294322-3 (replace 2294322-2) : Table Elevation Linear Actuator, Zeus thru H16 Table. See long description for details. Part #: 2294322-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Actiator, Linear, Zeus thru H16 Table, Pinned, R4 Ball Screw Fillet FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Per ECR2079949, 212008898, 5/19/09. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 2294322-3 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2079949 Part# 46-296561P1 converts to Part #: 2294322-2 in FEMC Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Actuator, Linear, Zeus thru H16 Table, Pinned FRU: Yes Status: Orderable Parts Offered as: Current: 2294322-2 Harvested: Not Available Exchange: Not Available Comments: Added per ECR ms13452 041210
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.