GE Healthcare · CTHiSpeed
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
System time or date is incorrect. See Long Description Note: Because the date command reads the Software Clock, and the software clock is an interupt driven clock, it can vary from the Bios clock, or hardware clock by a significant amount depending on system usage. If the system has not been rebooted for more than 12 hours it is better to reboot the system and then check the time before determining that adjustment to time is needed. A reboot may allow the system to adjust itself.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 6.3.1 for HiSpeed CT/i with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.3.1 6.3.1 for HiSpeed CT/i with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.3.1 6.3.1 for HiSpeed CT/i with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.3.1 Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2276664-3 6.3.1M4 HiSpeed CT/I Class C Software 2276665-3 6.3.1M4 HiSpeed CT/I Class M Software 2276666-3 2276663-3 6.3.1M4 HiSpeed CT/I App SW 6.3.1 Rhapsode Product 6.3.1 10/29/01 Only for Systems with Remote Tilt Option
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required