Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2601010715001826000100750042

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 260101071 The DCB has reset unexpectedly See Long Description for full error Error: 260101071 Host: orp Process: orp_subsys File: rcib_node_init.c Line: 376 Wed Jan 24 11:25:20 2007 Host : On-Board Rotating Processor Ermes # : 260101071 Exception Class : Secondary Severity : Pri/Soft File : rcib_node_init.c Line# : 376 Function : No System Function Reported Scan Type : None/Unknown Scan: 50018/1/0 The DCB has reset unexpectedly. The ORP has successfully reconnected to sub-system. Reason for reset as reported by CPU: External signal source. CPU reset register: 0x80 Node status: 0x382 State flags: 0x9800e82

Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the DCB See LD for more details Wed Nov 22 05:44:35 2006 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 50042/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 05:44:35:070 Log Series: 80 CAN serial link error. An CC error was detected when sending a message to the DCB. Message code: 1004 Node ID: 51 Object ID: 0 Send status: 5

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.