Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2600068365000026000682318729

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Gantry rotates but system will not scan. Gantry rotates but system will not scan

Error message

Error: 260006836 The Encoder and the Drive MUST be Sync'd/Referenced. See the LD for the complete error code Tue Nov 16 14:20:36 2004 Error: 260006836 Host: STC Process: Axial File: amc_exec.c Line: 2132 Function: AXIAL CONTROL : axial control Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:20:36:955 Log Series: 69 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. The Encoder and the Drive MUST be Sync'd/Referenced. (Cur State: 10 Cmd: 0x170a0048) HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Error: 260006823 Gantry (Axial) could NOT find the ''C'' pulse. See Long Description for full error Tue Jan 17 16:51:34 2006 Error: 260006823 Host: STC Process: Axial File: amc_exec.c Line: 3248 Function: AXIAL CONTROL : axial control Scan: 18729/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 16:51:34:165 Log Series: 144 Gantry (Axial) could NOT find the ''C'' pulse. (AMC State:4 Cmd:0x150c0008 Cmd Vel: 0.008 sec/rev). HW/SW Error: Retry Rx to verify probable Gantry Hardware problem.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues