Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011822650000260118283260118137

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118226 (60-0504 Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. see Long Description for full error Wed Oct 11 16:27:45 2006 Error: 260118226 orp orp_subsys gen_exec.c 2376 Wed Oct 11 16:27:43 2006 Host : Jedi Ermes # : 260118226 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Static Scan: 50000/1/1 (60-0504)Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. Z6,Z7= 0h,0h

Error: 260118283 (25-0593) LVPS 400: Power Suppy output set to off. see long description Tue Jul 1 17:20:59 2008 Error: 260118283 Host: orp Process: orp_subsys File: gen_exec.c Line: 2414 Tue Jul 1 17:20:58 2008 Host : Jedi Ermes # : 260118283 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 (25-0593)LVPS 400: Power Suppy output set to off. 3 over/under voltage FW errors in the same LVPS power supply output. Ext15V=0, Ext160V=1, Can15V=2, Gate24V=3, Main_inv=4, Heater1=5, Heater2=6,

Er

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2324118-3 VCT-JH4 Auxiliary Equipped. Aux Box See Long Description for details Part#:2324118-3 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT-JH4 AUXILIARY EQUIPED FRU: Yes Status: Orderable Comments: Set as Critical Part for Product CT_VCT, ms04278, 05Oct28 Added per VCT RSPL, ms13452 041205 Picture below
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required