Revolution XRd Unit will error 6C00 and/or 6D00 in digital mode after first exposure.

Revolution XRd Unit will error 6C00 and 6D00 in digital mode after first exposure.

Symptom

Revolution XRd Unit will error 6C00 and/or 6D00 in digital mode after first exposure. See long description for details relating to issue Consider the following scenario: ADS software reloaded Backup made to ADS hard drive and floppy After this, at some later date: IDC software reloaded IDC database restored from ADS hard drive After the system is rebooted and an acquisition attempted, a 6C00 "Error Status Received in IDC Prep Reply" error message (HW Status 01) is displayed on the Advantx console. If a "telnet idc" session is open when attempting the exposure, the following messages may appear: [CFileBadpixelList::isExistFile] file <BPLIST.LST> is not exist. IDC-ACQUISITION(2):IDC_PREP_STATE Apollo.uponPrepRecord -- sending to TITAN prepResp (1) --> PREP_WARNING

Error message

6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply)

Applies to

Models
Revolution, Advantx
Product
Revolution XRd
Area
X-ray R&F

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required