GE Healthcare · CTLightSpeed1607916071
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Old exam not in database keeps showing up when trying to edit patient data for another current exam. See Long Description. Old exam not in database keeps showing up when trying to edit patient data for any current exams Exam 1886 would also want to be edited when choosing any exam. All queues recon,archive, filming, network, and auto film are empty. Exam 1886 was not in /usr/g/data_management or under Retro Recon.
Error message
Error: 3015 Send message failed :QRS.9301 See long description for full error Tue Nov 27 10:38:09 2007 Error: 3015 Host: tediivct Process: netserver File: netproc.c Line: 357 Send message failed :QRS.9301 Error: 16079 [EPD] One non-CT/MR image type found: image skipped See Long Description for full error Thu Aug 20 18:34:30 2009 Error: 16079 Host: koict Process: editPatientData File: Line: 0 [EPD] One non-CT/MR image type found: image skipped Error: 16071 [EPD] Detected missing images See Long Description for full error Thu Aug 20 18:35:16 2009 Error: 16071 Host: koict Process: editPatientData File: Line: 0 [EPD] Detected missing images
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 08BW44.1 Software for all BrightSpeed Systems with AIO Console and all RT16 Systems with GOC4 Console See Long Description 08BW44.1 Software is used on all BrightSpeed Systems with AIO Consoles and all RT16 Systems with GOC4 Consoles. Software collector called “08BW44.X” p/n 5328867 contains the following items: 5323837 - LightSpeed Xtream Console Operating System Version - GEMS Linux 6.3.8 5323838 - LightSpeed Applications Software DVD Version 08BW44.X 5328828 - Xtream Serial-Over-LAN Service Software Version - 2.0 Supports: Westville or Jarrell DARC/VDARC 5323839 - Class C 5323840 - Class M NOTE: Class C/M install CD are not shipped with system. NOTE: The Xtream Serial-Over-Lan Service Software is not used during the LFC unless instructed via a pop-up message box "SOL Service - ...run the Serial Over LAN Service CD...". This CD is loaded directly into the DARC Node. Then press the reset (or cycle power) at the front panel of the DARC Node as required. This CD will auto-eject when the SOL load process is complete. NOTE: The Xtream Serial-Over-Lan CD is not used for All-In-One Console.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.