GE Healthcare · X-ray R&FPrecision10900021090004
Long and Vert Can Ampifier Node Guarding Errors See long description for full details Precision 500D Aretha ve…
Precision 500D Aretha Long and Vert Can Ampifier Node Guarding Errors
Symptom
Long and Vert Can Ampifier Node Guarding Errors See long description for full details Precision 500D Aretha version IDD Vertical Amplifier Node Guarding Error 1090002 and IDD Longitudinal Amplifier Node Guarding Error 1090004. Vertical Can Amplifier A23 has Red LED lit and Long Can Amplifier has no LEDs lit.
Error message
Error 1090002 and 1090004 See long description for description of error Error 1090002 - IDD Vertical Amplifier Node Guarding Error IDD Power Assist Vertical CAN amp initialization failed or is not responding. Indicates a CAN device or bus problem. Refer to the CAN bus block diagram. - If a single device is reporting an error, check that device, its power, and its cabling/connections. - If multiple device are reporting errors, check common cabling/connections and common power. For example, if the CAN bridge, Pos I/O, and CAN amps in the Pos cab are reporting errors, check the external CAN cable from Sys Cab to Pos Cab and check for power to those devices in the Pos Cab. - If all CAN devices are reporting errors, check common cabling/connections. Also check/replace the System Controller. Error 1090004 - IDD Longitudinal Amplifier Node Guarding Error IDD Power Assist Longitudinal CAN amp initialization failed or is not responding. Indicates a CAN device or bus problem. Refer to the CAN bus block diagram. - If a single device is reporting an error, check that device, its power, and its cabling/connections. - If multiple device are reporting errors, check common cabling/connections and
Applies to
- Models
- Precision
- Component
- Precision 500D Positioner Cabinet
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug
- Precision 500D Aurora, RCIM E-stop circuit is difficult to troubleshoot because there is no overall interconnect schematic from the RCIM operator console to the Corona 2 board in the System Cabinet.
- Precision 500D Aurora system with Magic PC.
- P500D Siemens Collimator lamp voltage adjustment instructions missing due to documentation issues.