GE Healthcare · CTLightSpeedVCT260132907260132912260132913260132910260122003
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Customer sees "Detector Temperature Is High. Image quality may be affected". Customer sees Detector Temperature Is High. Image quality may be affected
Error message
Error: 260132907 Rotating side (Plenum) CFC overtemperature condition detected. See Long Description for complete error Fri Feb 24 11:39:33 2006 Error: 260132907 orp orp_subsys ThermalManager.cxx 1588 Fri Feb 24 11:39:31 2006 Host : Das Ermes # : 260132907 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 1588 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Rotating side (Plenum) CFC overtemperature condition detected Error: 260132912 DAS Digital Module Temperature Out of Tolerance See Long Description for more details Wed Mar 22 00:23:01 2006 Error: 260132912 Host: orp Process: orp subsys File: ThermalManager.cxx Line: 2282 Wed Mar 22 00:23:00 2006 Host : Das Ermes # : 260132912 Exception Class : Secondary Severity : Sec/Soft File : ThermalManager.cxx Line# : 2282 Function : No System Function Reported Scan Type : None/Unknown Scan: 7/2/0 DAS Digital Module Temperature Out of Tolerance Digital Module: 28 Temperature: 5003 Limits: 3800, 5000 Error: 260132913 Plenum Plenum Filtered Average Temperature Out of Tolerance See long description for full error Error: 260132913 Plenum Plenum Filtered Average Temperature Out of To
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required