LightSpeed 5.X (60kW)

Symptom

Customer gets Error message 'DAS gain calibration has not been run since converter boards were changed' while running fast cals. Customer gets Error message 'DAS gain calibration has not been run since convertor boards were changed' while running fast cals.

Error message

Error: 244520 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error. Fri Apr 21 07:30:53 2006 Error: 244520 Host: oc01 Process: fwmgr File: HwManager.c Line: 1536 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error.

Error: 260116316 Communications are DOWN between subsystems See LD for full error Mon Jun 13 14:34:09 2005 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 904 Mon Jun 13 14:34:03 2005 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Primary File : lan_watchdog.c Line# : 904 Function : No System Function Reported Scan Type : None/Unknown Scan: 7754/5/0 Communications are DOWN between subsystems Or Sun Dec 17 08:26:45 2006 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 996 Sun Dec 17 08:26:39 2006 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Secondary File : lan_watchdog.c Line# : 996 Function : No System Function Reported Scan Type : None/Unknown Scan: 50119/1/0 Slip-Ring Communications a

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.