Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT24463426010100824457251004244540

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 244634 Subsystem State; - DCB:down -, CCB:up, TCB:up, JEDI:up, TGP/ORP:up DCB:DOWN See Long Description Fri Mar 9 08:35:05 2007 Error: 244634 Host: pct1 Process: fwmgr File: fwmgr_utils.c Line: 662 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:down CCB:up TCB:up JEDI:up TGP/ORP:up

Error: 260101008 The ORP failed to connect with the DCB sub-system in the allotted time. See LD Wed Jul 12 06:02:29 2006 Error: 260101008 Host: orp Process: orp_subsys File: rcib_node_init.c Line: 110 Wed Jul 12 06:02:30 2006 Host : On-Board Rotating Processor Ermes # : 260101008 Exception Class : Soft Severity : Pri/Soft File : rcib_node_init.c Line# : 110 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The ORP failed to connect with the DCB sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS

Error: 244572 UnExpected Event received sender:orp_subsys event:51004 time:0 See long description for full sample error message. Mon Aug 11 08:36:32 2008 Error: 244572 Host: h0150ct1 Process: fwmgr File: fwmgr_events.c Line: 9449 Function: No system function reported Scan Type:

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5130200 VCT DCB PWA See long description for details Part #: 5130200 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCT DCB PWA FRU: Yes Status: Orderable Parts Offered as: Current: 5130200 Harvested: Not Available Exchange: Not Available Comments: Added per VCT RSPL, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues