Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182405100642072920807184836

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182405 HW/SW error: Uncommanded elevation motion. See long description for full error Wed Mar 19 13:22:41 2008 Error: 182405 Host: ETC Process: Elevation File: unComMotion.c 1.28 11/5/92 Line: 170 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 10064/2/6 Type: Axial Exception Level: Pri/Soft Time: 13:22:42:215 Log Series: 54 HW/SW error: Uncommanded elevation motion. Current Position 20729 (.01 mm) Last Stationary Position 20807 (.01 mm) Position Tolerance 75 (.01 mm) Current Encoder cnts 3446 Last Stationary Encoder Cnts 3435

Error: 184836 motor controller detects uncommanded motion. Aborting any active scan Tue Aug 10 19:04:09 2004 Error: 184836 Host: ETC Process: Elevation File: elev scan.c 1.17 9/30/92 Line: 178 Function: No System Function Reported Scan: 4530/2/7 Type: Helical Exception Level: Pri/Most Severe Time: 19:04:08:905 Log Series: 376 motor controller detects uncommanded motion. Aborting any active scan

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269173 ELEVATION ENCODER ASM FOR H2 LCC -Zeus Table see Long Description for details and picture Part #: 2269173 Part Type: GP CONSUMABLES Part Weight: 3.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER ASM FOR H2 LCC FRU: Yes Status: Orderable Comments: FLS - ELEVATION ENCODER ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296633G3).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 WILL START DEPLOYMENT FROM ASIA ON FW12, 02. MS30544 1/31/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.