Symptom
During reboot, System stops with message: Press F1 to continue. During reboot of system the following message is posted. Error "512-chassis fan not detected" Press F1 to continue After pressing F1 system boots normal. During reboot of system the following message is posted. Error "512-chassis fan not detected" Presse F1 to continue After pressing F1 system boots normally.
Error message
Console exhibiting GRUB NOT LOADING with Error Codes 16 and 25 console exhibiting GRUB NOT LOADING with Error Codes 16 and 25
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5183547-38 HP8400 with 4GB Memory See Long Discription for Details Part #: 5183547-38 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: GEHC MI and CT Integrated Buy Level PC Workstation. HP8400 with 4GB Memory FRU: Yes Status: Orderable Comments: Per ECR 2050682 & GOC6 RSPL, 212008898, 7/14/08. ** FE ASSIST ** This part is not qualified for PET systems, CT only This part requires a Software Load from Cold LFC during replacement. For all CT and RT Products, verify you have the latest software and Service Pack for your system before trying to install this part. The software versions listed below are a reference, refer to the latest revision of the software Load Cold Document for your system for more information. For 4,8,16 slice CT HPower and RT4 slice systems, order 07MW11.10, For 4,8,16 slice CT Brightspeed systems, order 07BW08.4 For 16 slice RT systems, 08BW20.4 For 32, 64 slice GOC5 based systems, order 07MW18.4, For 32, 64 slice GOC6 based systems, order 08MW33.2 Future versions of the software releases noted above will are also compatible with this part. (212001261 10/21/08)
- Area
- CT
Full cause & step-by-step resolution