GE Healthcare · CTHiSpeed1854016193152125380116016176001
HSA RP 2.X HiSpeed Advantage
Symptom
Scanner Hardware Stopped Scan WITH View Length Error See Long Description for possible error Scanner Hardware Stopped Scan DURING Scan data acquisition WITH Error: 185401 slip-ring view length errors 1 and Error: 253801 View Length Error
Error message
Error: 185401 slip-ring view length errors 1 See Long Description for full error Host: SBC0 Proc: fwmgr Error: 185401 Exception Level: Pri/Soft SARQ statistics: total views 6151, retried subpackets 2 slip-ring view length errors 1 DAS view length error 0, DAS parity error 0 Taxi Link violations -61931521 Error: 253801 View Length Error. The size of the views received were incorrect See Long Description for complete error Mon Dec 6 12:12:06 2004 Error: 253801 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_err_handler Line: 2355 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 176001 FEP DSP 0: View Length Error. The size of the views received were incorrect (HDW status: 0x4006b7) (Data Type: 0x1). View Sequence Number: 6537. Possible causes: Slip Ring buffer terminator boards, SCOM board, or FEP board
Applies to
- Models
- HiSpeed
- Product
- HSA RP 2.X HiSpeed Advantage HiSpeed Advantage RP 2.X HiSpeed Advantage RP 2.X FMI 25247 HSA/RP 2.5 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.