Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1824271203314660

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Table elevation moves tiny increments and stops. Table elevation moves tiny increments and stops

Error message

Error: 182427 Gantry Back / Forward Touch Sensor Error See Long Description for complete Error Error: 182427 Gantry Tilt Back or Gantry Tilt Forward Sensor Error Error: 182427 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 125 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 12033/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 14:58:23:525 Log Series: 344 The Gantry BACK Touch Sensor Error : Console Pushbutton Remote Tilt is DISABLED. Use Gantry Controls to Tilt the Gantry. Current Position 14660 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down) (Elev Cmd: 1 TouchSensor: 0x8000). Same Error for the Forward Touch sensor

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required