GE Healthcare · CTLightSpeed2600050031818850152230006001182119
LightSpeed 4.X - Octane
Symptom
SRU has gone down. SRU has gone down
Error message
Error: 260005003 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Thu Apr 29 06:05:11 2004 Error: 260005003 Host: fitzipct Process: procICE File: iceCpu.cxx Line: 343 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Last ping was at Sun May 30 22:21:46 2032 with seq num 1818850152 iceConsole.log: Initializing System Memory (DRAM)... System Memory: 128MB, ECC Enabled (ECC-Memory Detected) PPC1-Bug Error: 230006001 Process presumed down: vxmonitor-post Thu Jun 22 14:57:42 2006 Error: 230006001 Host: ls01 Process: procICE File: iceCpu.cxx Line: 360 Process presumed down: vxmonitor-post Error: 182119 LAN Watchdog TIMEOUT for the SBC. See Long Description for more details Sun Sep 17 10:24:42 2006 Error: 182119 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 556 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:24:20:560 Log Series: 37 LAN Watchdog TIMEOUT for the SBC. HW/SW Error: Did not receive LAN watchdog message.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2260421 replaces 2260421-2 VME console power supply. Used on CTi - 4.x H3 GOC1 and GOC2 Operators Console. See Long Description for Details. 2260421 replaces 2260421-2 Part #: 2260421 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VME CONSOLE POWER SUPPLY ASSEMBLY FRU: Yes Status: Orderable Part number 2260421-2 is a comsumable assembly at EOL with 45 per year comsumsion. Part number 2260421 is the same base assembly with an added bracket and is a repairable. Do to the EOL issue and low comsumption of 45 per year we will eliminate p/n 2260421-2 and replace it with p/n 2260421 to have one repairable part. This will eliminate the need to stock two parts and create a new repair process.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.