Pular para o conteúdo principal
GE Healthcare · CT200501546200501547200001821200285067

Lightspeed 1.X - Octane

Symptom

SmartPrep failed when pause selected. SmartPrep failed when pause selected

Error message

Error: 200501546 Scanner software returned status -1 See Long Description for complete error Thu May 28 17:15:23 2009 Error: 200501546 Host: PRESTCT Process: scanRx File: uifIpcMgrScanRx.c Line: 4512 Function: Data Acquisition : OC Processing Scanner software returned status -1

Error: 200501547 Process: scanRx A Communication Error has occurred in uifIpcSendRecvSmartPrepStartUp Wed Dec 7 13:38:30 2005 Error: 200501547 Host: sect092 Process: scanRx File: uifSmartPrepIf.c Line: 187 Function: Data Acquisition : OC Processing A Communication Error has occurred in uifIpcSendRecvSmartPrepStartUp with seq_num = 1200 and status = 200001821

Error: 200285067 DisplayMgrControl::displayMgrTimerExec() process smartPrep has failed to establish communication with the manager. See long description Fri Feb 26 10:47:13 2010 Error: 200285067 Host: gcct Process: examRxDisplayMgr File: disMgrControl.cxx Line: 2103 DisplayMgrControl::displayMgrTimerExec() process smartPrep has failed to establish communication with the manager. This process will be terminated and restarted by the manager.

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
10.5_2.8.2I_H1.3M4 for LightSpeed 1.X (QXi) with Console Assembly (2180551) (Console with O2 Computer) or (2180551-2) (Console without O2 Computer) - showprods returns LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include(but are not limited to) the following:Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed Apps ver 10.5_2.8.2I_H1.3M4 Released via FMI 25269 The features introduced with this software include (but are not limited to) the following: Thin Twin Axial Mode (2 x .63 mm) Multi-Destination DICOM Print Capability Latest release of Iniste Interactive (IIP 2.1) Advanced Artifact Reduction (AAR) Web Browser Based Common Service Desktop - Phase II Includes a new HomePage New Log Viewer Separated Class A, C, and M software Beam-Tracking Improvement LightSpeed QX/i Rev 1.3 System Software Collector 2260334-5 consisting of 1.3 Software CD-ROM Collector 2258825-7 and Software Rating Plate 2240213-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240215-2 1.000 PC QX/I IRIX 65 VER 6.8.A DI 2240216-2 1.000 PC QX/I IRIX 65U VER 6U4.5.A 2258610-6 1.000 PC LIGHTSPEED QX/I APPS SFTW
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues