Pular para o conteúdo principal
GE Healthcare · CTHiSpeed

HiSpeed CT/i 4.X with Indigo and HSA Gantry

Symptom

The Data Collection Sub System has exceeded test limits for image quality, please contact service reported on user screen. The Data Collection Sub System has exceeded test limits for image quality, please contact service reported on user screen

Error message

Error: 0 Smart Trend: Limit Check failed for Air Scan See Long Description for full error Fri Mar 11 07:56:20 2005 Error: 0 Host: bhicct1 Process: dasgain File: dasGainOperations.c Line: 1390 Smart Trend: Limit Check failed for Air Scan, Short Term Ratio, Trend Data Look for details in error file : /usr/g/insite/ProDiags/qsa/results/dasgain_air_st_lp.err The error file is REMOVED after the next Fast Cal. After you run resetBaseline, the following normal message will be reported in the gesyslog: Fri Mar 11 08:00:38 2005 Error: 0 Host: UNKNOWN Process: UNKNOWN File: dstJob.c Line: 485 Base Line Ratio files not present. Creating them

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2102234 HP-DAS CONVERTOR BOARD See long description for details Part #: 2102234 Part Type: GP REPAIRABLES Part Weight: 1.36 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HP-DAS CONVERTER BOARD FRU: Yes Status: Orderable Comments: FLS - HP DAS 47 PIECES PER SYSTEM T3202GX HP-DAS CONVERTER BOARD GELS - "CAN BE USED ONLY ON HP-DAS, NOT ON E-DAS (FOR E-DAS, 46-183027P128IS THE FILTER BD, AND 46-183027P129 IS THE ADC BD) SP/JMG9/95 47 BOARDS PER SYSTEM. SP/JMG9/95 KANBAN CSS F00",PCB HP CONVERTER
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues