Pular para o conteúdo principal
GE Healthcare · CTHiSpeedLightSpeed185122182902

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Table will not move. Table will not move

Error message

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Error: 182902 Emergency Off button has been pressed Sun Sep 5 16:04:39 2004 Error: 182902 Suite: RP21 Host: ETC Process: Table Controls File: tc emergency.c 1.6 2/16/99 Method: No Method line: 89 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 4016 Log Series: 22 Emergency Off button has been pressed

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-297698P3 27" HIGH PROFILE Tape switch See the LD for tape switch jpg Part #: 46-297698P3 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 27" HIGH PROFILE BEAD, RIBBON SWITCH FRU: Yes Status: Orderable TABLE TOP COVER TAPE SWITCH. FRU = 2. *P3 WILL REPLACE THE P2 WHEN P2 STOCK IS DEPLETED. FE: IF YOU ARE REPLACING 46-297698P2 YOU MUST ORDER BOTH 2216522 AND 46-297698P3
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.