Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1832571876318310710785

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits

Error message

Error: 183257 KV board tube spit counter = xx See Long Description for complete error 183257 KV board tube spit counter = 1 Tue Nov 30 15:38:17 2004 Error: 183257 Suite: CT01 Host: OBC Process: Control File: gen_abort.c 1.33 Method: No Method Line: 716 Function: XRAY GENERATION : KV Loop Scan: 18763/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:38:16:115 Log Series: 387 KV board tube spit counter = 1

Error: 183107 KV fault detected: Anode shoot-thru. See long description for complete error Thu Mar 10 12:46:26 2005 Error: 183107 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 239 Function: XRAY GENERATION : KV Loop Scan: 10785/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 12:46:26:065 Log Series: 608 KV fault detected: Anode shoot-thru. Fault latch address = ffcff3 hex, Bit = D2

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues