Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT290014004

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner fails to boot to applications, Unable to verify DARC system startup. The system will not boot up to applications with a pink attention box: Unable to verify system functionality on startup. Please power cycle the console. If the system fails to startup after the power cycle, then please call GE Service. FE or OLE sees that they can telnet to the DARC but they cannot ping or rsh with the DARC. If the DARC is observed while booting, Drive X adapter X predicting future failure Press any key to continue is seen.

Error message

Do Not Use - use one of the other Error: 290014004 Unable to verify DARC system startup See Long description for complete error NOTE: This is error does not mean you need to replace the darc. Troubleshooting is necessary This error comes in different contexts per the below Mon Dec 12 08:00:11 2005 Error: 290014004 Host: ct99 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 172 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for ping darc response (exit status: 8) Fri Dec 8 14:42:12 2006 Error: 290014004 Host: kghcateqxi Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - scan disk array is non-recoverable (exit status: 22) Fri Mar 16 09:13:03 2007 Error: 290014004 Host: ctflsh Process: gre-verify-darc File: gre-verify-darc.cpp Line: 179 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting for rsh darc response (exit status: 9) Fri Mar 14 05:49:37 2008 Error: 290014004 Host: nvgect01 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 180 Unable to verify DARC system startup. /usr/g/bin/gre-verify-darc: failure - waiting

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Software
08MW33.2 for GOC6 LightSpeed VCT (64 Slice Only) 11/17/2008 Inital M3 release. This software is now obsolete, this site should be upgraded to 11HW12.5 software released via FMI 25427 08MW33.2 for GOC6 LightSpeed VCT (64 Slice Only) 11/17/2008 Inital M3 release. LightSpeed 7.1 ( 08MW33.2 ) Software Release supports GOC6 consoles used on VCT 64, VCT XT & VCT Std. - LightSpeed 7.1 (08MW33.2)Software for GOC6 Collector 5318369 Rev 1 Includes OS and Application CD's - Operating System Linux 5.2.13 Software CD 5318646 Rev 1 - Application LightSpeed VCT GOC6 DVD 5317935 Rev 1 - Class C LightSpeed 7.1 5318761 Rev 1 - Class M LightSpeed 7.1 5317906 Rev 1 An On-site/Remote identification aid . (Example: check_config) Result from check_config or swhwinfo 08MW33.2 V40_PS_HD64_G_GTM 08MW33.2 V40_PS_HD64_G_GTL Please create an ITRAK report for any software related issues
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required