GE Healthcare · CTLightSpeedVCT2601182832601182755010826011826050068
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118283 (25-0593) LVPS 400: Power Suppy output set to off. see long description Tue Jul 1 17:20:59 2008 Error: 260118283 Host: orp Process: orp_subsys File: gen_exec.c Line: 2414 Tue Jul 1 17:20:58 2008 Host : Jedi Ermes # : 260118283 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 (25-0593)LVPS 400: Power Suppy output set to off. 3 over/under voltage FW errors in the same LVPS power supply output. Ext15V=0, Ext160V=1, Can15V=2, Gate24V=3, Main_inv=4, Heater1=5, Heater2=6, Error: 260118275 (110-0903) Tube casing temperature has exceeded 92 degree C See LD for more details Sun Nov 6 12:44:33 2005 Error: 260118275 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Sun Nov 6 12:44:32 2005 Host : Jedi Ermes # : 260118275 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50108/1/3 (110-0903)Tube casing temperature has exceeded 92 degree C.^C Z6,Z7= 0h,0h Error: 260118260 ( 80-1422 ) Overpress or Blown fuse Error: 260118260 ( 80-1422 )Overpress or Blown fuse Wed Aug 17 17:37:03
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5151365 Hercules Heat Exchanger 48 Volt Power Supply Assembly with Cover See LD for more information Part #: 5151365 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Hercules Heat Exchanger Power Supply Assembly with Cover, part numbers 2306440 and 2298317 FRU: Yes Status: Orderable Parts Offered as: Current: 5151365 Harvested: Not Available Exchange: Not Available Comments: Added per ECR 2019074, this part is Power supply for VCT and HPower Heat Exchangers, ms13452 051120 POWER SUPPLY MANUFACTURED BY TOTAL POWER INTERNATIONAL INC. 418 BRIDGE ST. LOWELL, MA. 01850 350 WATT SINGLE OUTPUT SWITCHING POWER SUPPLY, UNIVERSAL INPUT, U- BRACKET 48 VOLT, 7.3 AMP PART NUMBER TPS350-14
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.