Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2300230522446531163411915

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Message on screen "A system communication failure has occurred which will prevent scanning. Please contact GE service" A system communication failure has occurred which will prevent scanning. Please contact GE service

Error message

Error: 230023052 Failed in Interprocess Communication with Firmware. See long description for details Fri Apr 4 07:32:59 2008 Error: 230023052 Host: chhvct Process: dailyPrepRx File: DailyPrepShellMgr.c Line: 2618 Failed in Interprocess Communication with Firmware. Critical software failure. Cannot continue.

Error: 244653 Host Querying Detector Temperature Status See LD for complete error Sun Jun 25 13:08:53 2006 Error: 244653 Host: gect02 Process: fwmgr File: fwmgr_events.c Line: 11634 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Host Querying Detector Temperature Status: StateFlag:2 curTe mp:3180, leftTemp:3620 rightTemp:3730 Time:0 Fri Feb 25 07:51:11 2011 Error: 244653 Host: MT_CT2 Process: fwmgr File: fwmgr_events.c Line: 11915 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Host Querying Detector Temperature Status: StateFlag:5 curTemp: 0 , leftTemp: 0 rightTemp: 0 Time:0 <--- All temperatures reporting zero - Check for Blown F6 on Fuse Board

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.