Pular para o conteúdo principal
GE Healthcare · CTLightSpeed182825182872105252372485182826

LightSpeed 4.X - GRE

Symptom

Scanner hardware reset and flash download fails

Error message

Error: 182825 HW/SW error: Communication Problems with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182825 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 30 HW/SW error: Communication Problems with the GANTRY DISPLAY. Number of Communication Timeout errors: 0. Number of NACK's errors: 3. Number of DUART errors: 0. Numver of Checksum errors: 0.

Error: 182872 TNC Controller Received an Invalid Command. See LD for more details Error: 182872 TNC Controller Received an Invalid Command. Thu May 25 08:19:06 2006 Error: 182872 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 505 Function: No System Function Reported Scan: 10525/2/2 Type: Helical Exception Level: Pri/Soft Time: 08:19:06:480 Log Series: 140 TNC Controller Received an Invalid Command. Message Code: 7 Current State: 1. Board Number: 2372485 Serial Number: 0.

Error: 182826 Communication Error: Communication Packet with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182826 Host: ETC Proce

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2372485 ETC IF (Interface or TNC Bd.) Replaces 2229871 See long description for details 2372485 ETC IF (Interface or TNC Bd.) Replaces 2229871 Part #: 2372485 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: ETC INTERFACE W/ AZ60A FRU: Yes Status: Orderable Comments: FLS - ETC INTERFACE W/ AZ60A. ADDED PER PCN 220857. REPLACES P/N 2229871. USE UP EXISTING STK. MS20544 5/1/03. ONE USE LIGHTSPEED PLUS VC:611695-SCI, RAPID CITY, SD CONTROL PCB 46-264368G3
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.