GE Healthcare · CTLightSpeedVCT26011822650000260118220
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118226 (60-0504 Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. see Long Description for full error Wed Oct 11 16:27:45 2006 Error: 260118226 orp orp_subsys gen_exec.c 2376 Wed Oct 11 16:27:43 2006 Host : Jedi Ermes # : 260118226 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Static Scan: 50000/1/1 (60-0504)Inverter Gate Power Supply error. gate supply voltage has dropped below the level required to drive the IGBTs properly. Z6,Z7= 0h,0h Error: 260118220 (60-0325) DSP overload See Long Description Sun May 30 15:50:49 2010 Error: 260118220 Host: orp Process: orp_subsys Host : Jedi Ermes # : 260118220 Exception Class : Abort Severity : Primary Scan Type : Helical Scan: 9451/2/2 . (60-0325)DSP overload Code 60 Exposure Errors Error code: 60-0325H Message/explanation: DSP overload Default in kV regulation calculation Potential cause: Main supply Inverter current measurement PPC kV control board Inverter HV tank defective Check main DCBUS voltage on inverter Recommended action/ Troubleshooting guide: Check if transformer current
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2281950-3 or 2281950-4 : JH4 INVERTER WITH SHIELDS used on LS 5.X (80/100), LS 7.X See Long Description for Details 2281950-3 Can also use 2281950-4 JH4 INVERTER WITH SHIELDS used on LS 5.X (80/100), LS 7.X Part #: 2281950- 3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: added per ecr 2005924, as replacement for 2281950 and 2281950-2. ms13452 040923 Set as Critical Part for Product CT_VCT, ms04278, 05Oct14 - If you receive an inverter as FRU (P/n 2281950, 2281950-2, 2281950-3) that must be installed to replace a defective inverter on site, check first Part number of PPC kV control board in this FRU: - if it is labelled 2351200-11 rev 001 this board needs to be replaced with the kV control board that is installed in the defective inverter: PPC board replacement, please refer to PPC FRU replacement procedure in service manual. - if it is not labelled 2351200-11 rev 001, the board installed on FRU is good: no modification is needed. Shawn Fox 3-24-05 Part #: 2281950 -4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBRPart #: 2281950-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: Added per ECR 2018129, THIS PART IS BACKWARD COMPATIBLE REPLACEMENT FOR 2281950-3, -2, & NO DASH. MS13452 050919 Repairable: No Returnable: No Hazardous Material: N Description: JH4 INVERTER WITH SHIELDS FRU: Yes Status: Orderable Comments: Added per ECR 2018129, THIS PART IS BACKWARD COMPATIBLE REPLACEMENT FOR 2281950-3, -2, & NO DASH. MS13452 050919
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required