GE Healthcare · CTHiSpeed2300075132000027231071820000271826390
Scout images fail to reconstruct.
Symptom
Scout images fail to reconstruct. See long description for details Scout images fail to reconstruct. Rebooting, cycling console power, etc. does not help.
Error message
Error: 230007513 ICPROXY: checksum failed while dma'ing image, expected ..., calculated ... See Long Description for complete error. Error: 230007513 ICPROXY: checksum failed while dma'ing image, expected ..., calculated ... Tue Jun 20 17:16:30 2006 Error: 230007513 Host: ice-post Process: icproxy File: ic_Callbacks.cxx Line: 532 ICPROXY: checksum failed while dma'ing image, expected 0x89f395ca, calculated 0xdfa7108f , seq id 1024 Error: 200002723 Pixel data response : Erros status SYS_ERR See LD for Full error Error: 200002723 Pixel data response : Erros status SYS_ERR Tue Jun 20 17:16:30 2006 Error: 200002723 Host: ct Process: imagecreate File: ic_getpixeldata.c Line: 373 Image : 10718/1/1 Recon Sequence Id : 1024 Pixel data response : Erros status SYS_ERR Error: 200002718 Image create Pixel data request failed See the long description for the full error code Thu Jun 30 23:03:01 2005 Error: 200002718 Host: ip01 Process: imagecreate File: ic imagecreate req.c Line: 247 Image : 26390/2/3 Recon Sequence Id : 15 Image create Pixel data request failed
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) See FMI 25903 I Helios_H2_O2LESS_CLASSC_Product 09/21/2005 Helios ver 1104L.3 rel 1104L.3_H4.2CLASSC I Helios_CLASSMOC_Product 09/21/2005 CLASSMROOT ver 1204L.3 rel 1204L.3_H4.2CLASSM Class C is 2394897 Order the following P/N: 2401759 (6 pack) Class M is 2394898 Order the following P/N: 2401759 (6 pack) Software Collector 5115431 Recon Patch for GOC2 only 5123440
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.