Pular para o conteúdo principal
GE Healthcare · CTHiSpeed25419360612254157262142216658216656

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Recon self test failure on boot up. Recon self test failure on boot up

Error message

Error: 254193 IG0 VME Image Memory test failed. See long description for full error Sun Sep 27 16:32:45 2009 Error: 254193Suite: CT01 Host: SBC Process: Diag ExecFile: igDiagTests.c 1.10 Method: invokeIgDatapath0 Line: 757Function: Diagnostics No System Function ReportedTest Name: SBC IG0 Datapath Unique ID: 60612 Iteration: 1Exc. Level: Soft Time: 0 Error Type: Hardware IG0 VME Image Memory test failed. See previous error message(s) for more information.^CThe igVmeImageMemTest() function call failed.

Error: 254157 Image Memory 1 failed the read/write data verification test via the VMEbus See long description for full error Sun Sep 27 16:32:45 2009 Error: 254157 Suite: CT01 Host: SBC Process: Diag Exec File: igDiagTests.c 1.10 Method: invokeIgDatapath0 Line: 740 Function: Diagnostics No System Function Reported Test Name: SBC IG0 Datapath Unique ID: 60612 Iteration: 1 Exc. Level: Soft Time: 0 Error Type: HardwareIG0 Image Memory data error. Image Memory 1 failed the read/write data verification test via the VMEbus. Possible cause: IG board, backplane, SBCData: Exp=0xaaaaaaaa Act=0x55555555 Failed Bits=0xffffffffMatches=262142 Mismatches=2 Offset=0xc

Error: 216658 Image reconstruc

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2118899-2 RHAPSODE IMAGE GENERATOR Board (IG Board). Used in Lightspeed 1x, CTi 4x, CTi 6x systems. See LD for details Part #: 2118899-2 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: RHAPSODE IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - IMAGE GENERATOR BOARD USED ON RHAPSODE MS1356 BURR 12-95 REPLACED 2118899 PER PCN 189598. MS4636 2/6/96 PCN 189939 GELS - KANBAN CSS F00,PCB IMAGE GENERATOR
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues