GE Healthcare · CTLightSpeed2600010072600015145000426010101250060
LightSpeed 5.X (80/100kW)
Symptom
Scan hardware stopped scan with RCIB errors
Error message
Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5 Error: 260001514 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. See long description for full error Wed Apr 27 01:00:51 2005 Error: 260001514 Host: orp Process: orp_subsys File: notiftsk.c Line: 1368 Running: System Reset Wed Apr 27 01:00:51 2005 Host : Das Ermes # : 260001514 Exception Class : Abort Severity : Secondary File : notiftsk.c Line# : 1368 Function : No System Function Reported Scan Type : None/Unknown Scan: 50004/1/0 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. (Timeout:500 msecs). Error: 260101012 The Rotating Controller Interface Bus (RCIB) fault
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2318704 DCB MDAS 16 slice See LD for details Part #: 2318704 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -14549.85 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 16 SLICE DCB MDAS FRU: Yes Status: Orderable Parts Offered as: Current: 2318704 Harvested: Not Available Exchange: Not Available Comments: FLS - 16 SLICE DCB MDAS ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required