GE Healthcare · CTHiSpeed200002738200285080
The autoview viewport no longer works and unable to display some images from previous exams
Symptom
The autoview viewport on Exam Rx desktop no longer works and also unable to display some images from previous exams listed on the system disk.
Error message
Error: 200002738, Creation of hard link to Image file for autoview failed See Long Description for complete error Tue May 23 10:37:48 2006 Error: 200002738 Host: ct01 Process: imagecreate File: ic_imagecreate_req.c Line: 1361 Image : 2928/3/17 Recon Sequence Id : 201 Creation of hard link to Image file for autoview failed Autoview file path : Varies Source file path :Varies Open file name : Varies Directory stat status : 0 File stat status : 0 Errno : OS error : 2 --- No such file or directory Error : 200285080 : An invalid autoview path was received from the imageAvailable event. See Long description for full error. Sat Jul 4 12:42:54 2009 Error: 200285080 Host: ct02 Process: examRxDisplayMgr File: /bat_pool/work22830/commondisplay/src/cxx Line: 1152 validPath() An invalid autoview path was received from the imageAvailable event. autoview_path = |/usr/g/sdc_image_pool/ct_autoview/exam4250_series2/E4250U0_S2U0_I184U0.1246725737_287678| stat() returns with status -1, errno = The manager will not attempt to remove or pass this event to the avViewer and/or xrViewer .
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) 103L.5_HSQXIM4 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) See FMI 25903 I Helios_H2_O2LESS_CLASSC_Product 09/21/2005 Helios ver 1104L.3 rel 1104L.3_H4.2CLASSC I Helios_CLASSMOC_Product 09/21/2005 CLASSMROOT ver 1204L.3 rel 1204L.3_H4.2CLASSM Class C is 2394897 Order the following P/N: 2401759 (6 pack) Class M is 2394898 Order the following P/N: 2401759 (6 pack) Software Collector 5115431 Recon Patch for GOC2 only 5123440
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.