GE Healthcare · CTLightSpeed200001830230023002244640244634244540
LightSpeed 4.X - GRE
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 200001830 IPC failure caused by cupIpcConnect from scan_mgr to dataacq Open Long Description Fri Jul 1 16:51:14 2005 200001830 4 ct02 scanmgr ecommL_func.c 106 IPC failure caused by cupIpcConnect from scan_mgr to dataacq. Error : 230023002 Scanner Hardware is not ready or unavailable See LD for more info Fri Jun 10 08:22:33 2005 230023002 4 cr_ct_01 scanUtilsRx DetailedCalUtils.c 1160 FAILED to execute the Scan Request. Possible Reasons: Scanner Hardware is not ready or unavailable. Please verify and try again. *** Laser indicator flashing on the Gantry Display board – this indicates the system has lost communication with the ORP board. *** Additional error info: Error: 244640 HARD RESET through serial connection failed. some controllers(TGP, ORP) do not response to ping. Error: 244634 Immediate Request cannot be honored since the firmware is in the DOWN state Fri Jun 10 08:22:13 2005 230023002 host: cr_ct_01 process: scanUtilsRx DetailedCalUtils.c 1160 FAILED to execute the Scan Request. Possible Reasons: Scanner Hardware is not ready or unavailable. Please verify and try again. Fri Jun 10 08:22:33 2005 230023002 4 cr_ct_01 scanUtilsRx DetailedCalUtils.c 1160 FAILED to exe
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5144058-500 Upgrade kit - GOC3 to GOC4. DARC2 is not included. Order proper DARC2 separately - REFER to SN11063012 See Long Description for Details. 5144058-500 Upgrade Kit GOC3 to GOC4 - without DARC2 Part #: 5144058-500 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Service Kit - SDDA and GOC3 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL FRU: Yes Status: Orderable Comments: ** FE ASSIST ** This part number no longer contains a DARC2 it must be ordered in addition per Your system type. REFER to SN11063012a(July 26 2011 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.