GE Healthcare · CTLightSpeed18221120525750009085218416539142
LightSpeed 4.X - GRE
Symptom
Grinding noise coming from the table while driving cradle in or out. Table may move slower than normal. Customer may hear a grinding noise coming from the table while driving cradle in or out. Table may move slower than normal.
Error message
Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Aug 29 08:09:30 2006 Error: 182211 Host: ETC Process: Cradle File: cmcLm628IsrTsk.c 1.19 10/16/9 Line: 185 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 20525/2/1 Type: Helical Exception Level: Pri/Soft Time: 08:09:30:260 Log Series: 595 HW error: Velocity error detected by hardware (LM628 excessive position error). Expected Velocity 75000 (.001 mm/sec) Actual Velocity 0 (.01 mm/sec) Pos Err Tol 2540 (.01 mm) Cur Pos 90852 (.01 mm) Cradle Dir: 1 (0 = Out 1 = In [Gantry]). Error: 184165 Cradle was not ready. See long description Mon Jul 7 08:22:54 2008 Error: 184165 Host: STC Process: Scan Control File: scHdwRdy.c 1.36 Line: 77 Function: No System Function Reported Scan: 39142/3/1 Type: Helical Exception Level: Sec/Soft Time: 08:22:55:265 Log Series: 819 Cradle was not ready.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269169 CRADLE DRIVE ASM FOR CATHAY TABLE. See Long Description for details Part #: 2269169 Part Type: GP REPAIRABLES Part Weight: 14.23 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASM FOR CATHAY TABLE FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASM FOR CATHAY TABLE. P/N 2269169 REPLACES P/N 46-296235G2 PER PCN 220590USE FIRST EXISTING STOCK. MS20544 ON 8/23/02.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required