GE Healthcare · CTLightSpeedVCT26000240350241
LightSpeed 1.X - GRE
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260002403 The Data Acquisition System Has Detected An Error In The +5V Converter Board Digital DAS Power Supply. See long descripton for a sample of complete error message. Mon Dec 10 01:59:42 2007 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 674 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50241/1/0 Type: None/Unknown Exception Level: Sec/Soft Time: 01:59:42:620 Log Series: 208 The Data Acquisition System Has Detected An Error In The +5V Converter Board Digital DAS Power Supply.^B Measured: 4749 mV, Expected range: 4750 to 5250 mV
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required