Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1832966500118310833003

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Stopped Scan with High Voltage errors

Error message

Error: 183296 Monitoring: (Measured 245.5 mA vs Ref 240 mA) See Long Description for the full error Error: 183296 Suite: CT01 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 529 Function: No System Function Reported Scan: 65001/8/38 Type: Axial Exception Level: Sec/Soft Time: 03:11:20:495 Log Series: 49 Selected Technic: 100 kV, 240 mA, 0 % Mod, Backup Timer 4324. VarmA: (Index: Start 0, Skip 0, Last 0) Last Cmd 240 mA. Monitoring: (Measured 245.5 mA vs Ref 240 mA) . 5 Total Samples, Tube Spit Ctr 0. (Rotor 0x0000ae04, Gen 0x9c19b5ff, Mon 0x00020000

Error: 183108 KV fault detected: Cathode shoot-thru. See long description for full error Sun Jun 5 23:17:05 2005 Error: 183108 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 248 Function: XRAY GENERATION : KV Loop Scan: 33003/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 23:17:05:545 Log Series: 424 KV fault detected: Cathode shoot-thru. Fault latch address = ffcff3 hex, Bit = D3

Applies to

Models
LightSpeed
Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2208852 - HP INVERTER RHAPSODE - used on CT/i, 1.x, 2.x, 3.x and 4.x systems. 2208852 HP INVERTER RHAPSODE 2208852 HP INVERTER RHAPSODE Part #: 2208852 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: HP INVERTER RHAPSODE. FRU: Yes Status: Orderable Comments: IF RHAPSODE SYSTEM CURRENTLY HAS T3202JK CAN USE 2208852 . THE LONGER BOLTS NEEDED FOR THE 2208852 INVERTER ARE NORMALLY SUPPLIED IN A BAG WITH THE INVERTER OR CAN BE ORDERED SEPARATELY UNDER P/N 46-208561P43
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.