Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230019002 The scan disk I/O layer failed to consume a response from the scan disk server. See LD for complete error Tue Nov 2 03:19:30 2004 Error: 230019002 Host: sgredct1 Process:scanfilemgr File: sdio proxy.c Line:174 The scan disk I/O layer failed to consume a response from the scan disk server. function:consume sequence:6708

Error: 245050 Error verifying that Data Acquisition is ready for scanning Open Long Description to see error examples Error: 245050 Error verifying the Data Acquisition is ready for scanning. More than one error text exists for this error below: Fri Jul 1 16:51:14 2005 Error: 245050 Host: ct02 Process: scanmgr File: start_scanning.cxx Line: 231 Function: System Control : Scan Control Scan Type: Not scanning Scan: 483/1/0 Scan Seq Id: 1 Error -1 verifying that Data Acquisition is ready for scanning Wed Mar 3 16:56:16 2010 Error: 245050 Host: uhctlspro Process: scanmgr File: start_scanning.cxx Line: 250 Function: System Control : Scan Control Scan Type: Not scanning Scan: 36505/1/0 Scan Seq Id: 0 Error 200001821 verifying that Data Acquisition is ready for scanning

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5159833 VDARC - Volume Data Acq Recon Control Computer, 2nd Edition (Jarrell). See Long Description for Details.. Part #: 5159833 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDARC - Volume Data Acq Recon Control Computer, 2nd Edition FRU: Yes Status: Orderable Parts Offered as: Current : 5159833 Harvested : Not Available Exchange : Not Available Comments: Per Jarrell RSPL, 7/17/06, ms11892. ** FE Assist ** Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better.(12/8/06 212001261
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.