Scanner Hardware Stopped Scan with open filament error in log. Scanner Hardware Stopped Scan with open filament error

Error message

Error: 260118185 (50-0204) Filament permanent open circuit See Long Description for complete error. Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118185 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (50-0204)Filament permanent open circuit. Issued by the heater board when an open has been detected and 3 restarts have been tried without success within 100 ms. ^C Z6,Z7= 97h,0h

Error: 260118120 (20-0255) Filament Open circuit Ermes: 260118120 (20-0255) Filament Open circuit Wed Mar 21 08:25:45 2007 Error: 260118120 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118120 Exception Class : Secondary Severity : Pri/Soft File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (20-0255)Filament open circuit(inverter1)(<3 times). ^C Z6,Z7= 97h,0h

Error: 260118192 (50-0211)Jedi detected open filament circuit. See Long description for full error Host: Jedi Ermes #: 260118192 Exception Class: Abort Severity: Primary File: gen_

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5128609-8 (replaces 2201185-3 - P9230TP) - HV Cable Cathode 2.3M - Used on Jedi HP 60 generator See Long Description for Details 5128609-8 Cathode HV Cable (2.3 meters long. Used on LightSpeed HP60 and BrightSpeed Systems) Part #: 5128609-8 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: HV CABLE FRU: Yes Status: Orderable Comments: **FE ASSIST** 2300mm long HV cable from Cathode to HV tank Negative (212001261 May 4 2011) added per to NPI#019477 by sso#305010201 in Mar 28, 2010 Old FRU Compatible No Comments added per to ECR2109209 by sso#305024364 in Nov 2, 2010 Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5128609-8 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Direct ECO Number ECR2109209 2201185-3 - P9230TP - HV Cable Cathode 2.3M - Used on Jedi HP 60 generator, Model # 2371333-2 Part #: 2201185-3 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9230TP HV CABLE CATHODE 2.3M FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution