GE Healthcare · CTLightSpeedVCT18242615403
LightSpeed 4.X - Octane
Symptom
Customer was not able to remote tilt gantry. In many cases , gantry will tilt from the Gantry Controls.
Error message
Error: 182426: Gantry Touch Sensor FAULT. Console Pushbutton Remote Tilt is DISABLED See Long Description for Full Error Tue Jan 10 17:42:00 2006 Error: 182426 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 97 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7855/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:41:59:940 Log Series: 197 The Gantry Touch Sensor FAULT. Console Pushbutton Remote Tilt is DISABLED. (Gantry Sensors - FORWARD Touch Sensor Error(s)). Gantry Controls are running in FAULT MODE. Current Position 15403 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down). (Elev Cmd: 1 TouchSensor: 0x5000).
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required