Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011631650119

LightSpeed 5.X (80/100kW)

Symptom

Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board

Error message

Error: 260116316 Communications are DOWN between subsystems See LD for full error Mon Jun 13 14:34:09 2005 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 904 Mon Jun 13 14:34:03 2005 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Primary File : lan_watchdog.c Line# : 904 Function : No System Function Reported Scan Type : None/Unknown Scan: 7754/5/0 Communications are DOWN between subsystems Or Sun Dec 17 08:26:45 2006 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 996 Sun Dec 17 08:26:39 2006 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Secondary File : lan_watchdog.c Line# : 996 Function : No System Function Reported Scan Type : None/Unknown Scan: 50119/1/0 Slip-Ring Communications are DOWN between subsystems

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5370000 - TGPHD BOARD ASSY - TGPU For HD, HP and VCT (Replaces the 5271005-2 & 2349697-2, -3, -4 and -5). See Long Description about backward compatability. Part #: 5271005-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPHD BOARD ASSY Positioning VCTHD FRU: Yes Status: Orderable Comments: Per ECR 2098435, 212008898, 30 April 2010. Old FRU Compatible Yes Comments Per ECR 2098435, 212008898, 21 Jul 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5271005-2 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2098435 2349697-5 is replacing the -4 version of this board. -3 versions are still in stock. If you are replacing parts in a VCT or HP gantry order the -3 until stock is depleted. ** FE ASSIST ** The 2349697-3 version is for VCT and all 5.x systems i.e. HP60 HP80- HP100's. The 2349697-4 version will works for CT750-HD and is backwards compatible but the -3 will not work on a CT750HD. (212001261 12/11/08) Part #: 2349697-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPU BD ASSY Positioning VCTHD FRU: Yes Status: Orderable Part #: 2349697-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TGPU BD ASSY Positioning HD FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The 2349697-4 is for CT750-HD and is backwards compatible to The 2349697-3 used on VCT and all 5.x systems i.e. HP60 HP80- HP100's. Do not use the 2349697-3 on CT750HD (212001261 12/11/08) TGPU 2349697-3 Part #: 2349697-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPU BD ASSY Positioning HP60 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, this is replacement for -2, and non dash versions. ms13452 050325
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues