Symptom
Scanning aborted during a scout scan with Cradle position is out of Limit See Long Discription for Details. Scanning aborted during a scout scan with Cradle position is out of Limit
Error message
12-1001-41 cradle out of limit. (tp) see long description scout scan error detected by tgp board (for earlier than system version 5.5) scout scan error: 41 code: 12-1001-41 name: cradle out of limit. (tp) class: error description: cradle position out of limit on scout scan cmd. potential cause / fru: 1. cradle potentiometer 2. ims potentiometer
Applies to
- Models
- HiSpeed, VCT
- Product
- CT/e Dual, CT ProSpeed AII/FII Irix or Linux HiSpeed CT/e Dual CT Scanner System see long Description for sales information Product Name: CT/e Dual CT Scanner System Irix A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) N/A Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- U0008AP ROPE See Long Description Part #: U0008AP Part Type: GP CONSUMABLES Part Weight: 0.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: ROPE FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.