GE Healthcare · CTLightSpeedVCT182417
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Table elevation will not reference, flashing eights will not reset
Error message
Error: 182417 Invalid Index Pulse with reference to the Interference switches Sun Sep 18 10:02:45 2005 Error: 182417 Host: ETC Process: Elevation File: emcMonitor.c 1.44 1/29/96 Line: 111 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:02:41:090 Log Series: 27 HW error: Invalid Index Pulse with reference to the Interference switches.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269173 ELEVATION ENCODER ASM FOR H2 LCC -Zeus Table see Long Description for details and picture Part #: 2269173 Part Type: GP CONSUMABLES Part Weight: 3.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER ASM FOR H2 LCC FRU: Yes Status: Orderable Comments: FLS - ELEVATION ENCODER ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296633G3).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 WILL START DEPLOYMENT FROM ASIA ON FW12, 02. MS30544 1/31/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required