GE Healthcare · CTLightSpeed2601163075100151000
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260116307 Could not send message to 101 or 201 from 101 or 201 (any combination) see long description for error details. Error: 260116307 Could not send message to 101 or 201 from 101 or 201. See examples below: Fri Aug 15 15:01:01 2008 Error: 260116307 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 513 Fri Aug 15 15:00:48 2008 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan_watchdog.c Line# : 513 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 201 from 201. ( TGP from TGP) Event Attempting to send 51001 System Error - 0 Sun Jul 5 13:49:14 2009 Error: 260116307 Host: tgp Process: tgp subsys File: lan watchdog.c Line: 568 Sun Jul 5 14:00:47 2009 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan watchdog.c Line# : 568 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 101 from 201. (ORP from TGP) Event Attempting to send 51000 Tue Dec 15 14:28:25 2009 Error: 260116307 Host: orp Process: orp subsys File: lan watchdog.c Line: 568 Tue Dec 15 14:24:2
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2106993-5 FUSE 20 AMPS 500 VOLTS .3 SECONDS See Long Description for details Part #: 2106993-5 Part Type: GP CONSUMABLES Part Weight: 0.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 20 AMPS 500 VOLTS .3 SECONDS FRU: Yes Status: Orderable Comments: FLS - 20AMPS 500V .3SECOND FUSE U/O COST REDUCED GENERATORS SCPU 65/80 MS12746 PER RSPL 5-97 GELS - MAJ LEAD-TIME LE 99/07/26 PA,FUSE 20A 500V .3 SECONDS
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required