Symptom
Scanner Hardware stopped with WARNING: DAS Temperature is nearing the Upper Limit.
Error message
Error: 260007428 WARNING: DAS Temperature is nearing the Upper Limit. See LD for more details Mon Aug 28 06:39:06 2006 Error: 260007428 Host: orp Process: orp_subsys File: ctemptst.c Line: 600 Mon Aug 28 06:39:04 2006 Host : Das Ermes # : 260007428 Exception Class : Soft Severity : Sec/Soft File : ctemptst.c Line# : 600 Function : No System Function Reported Scan Type : None/Unknown Scan: 50359/1/0 WARNING: DAS Temperature is nearing the Upper Limit. Highest temperature was seen is on Converter Board 23. Measured: 63.0 degrees C. Acceptable Range: 15.0 to 62.0 degrees C. See Following Error Message for Converter Card List
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5340738 (replaces 2359114) HPOWER GANTRY FAN AND HARNESS used on 5.x gantry See Long Description for details This fan for BrightSpeed, HP100, HP60 and RT Systems. Part #: 5340738 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: Per ECR 2091297, 212008898, 10/26/09. Part #: 2359114 <-- All GPRS stock is gone and this part is now obsolete Part Type: GP CONSUMABLES Part Weight: 8.82 LBR Repairable: No Returnable: No Hazardous Material: N Description: HPOWER GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: FLS - HPOWER GANTRY FAN AND HARNESS.ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19, 2003. MS20544 2/28/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required